2.2.3 Wiring the Benchtop Optica

2.2.3a
Input Power
The Optica operates with input power from 100 to 120 VAC at 4 amps, or from 200 to 240 VAC at 2.5 amps. It is
designed for a nominal 100, 115 or 230 VAC source. A switch on the rear panel selects the appropriate voltage range
(see Figure 9 below).
The 115 VAC setting operates over a range of 100 to 120 VAC
The 230 VAC setting operates over a range of 200 to 240 VAC
The Optica's voltage and frequency rating are listed on the rear panel.
2.2.3b
Sensors
Connect the dew point sensor cable to the 25-pin connector in
Connect the optional temperature sensor cable to the 9-pin connector in
sensor and other I/O wiring connects to the terminal block in
*Note: Depending on the model revision, the LAN connector is located in one of two places.
Optica™ Operator's Manual
Slot B
Slot A
Figure 9: Optica Benchtop Rear Panel
on the Optica's rear panel (see Figure 9 below).
on the rear panel. The optional pressure
Slot B
.