Figure 4-4
Clear this check
box before starting
calibration
Click here to start
calibration
Calibration Dialog
Clear Cal Memory Protect to Run self calibration
You cannot run self calibration if this box is checked.
Click Start, then follow the instructions on the screen.
4
You will be asked first to disconnect all channels and the Aux Output, then to connect Aux Out
to each channel and the auxiliary trigger in turn. After you complete each cable change, click
OK in the dialog box to continue the calibration.
The oscilloscope displays a Passed/Failed message as each calibration routine is completed.
After calibration has been completed, click to check the Cal Memory Protect box to
5
protect the calibration memory. Then click Close.
If calibration fails
Go to chapter 5 "Troubleshooting."
Chapter 4: Calibrating and Adjusting
To run the self calibration
4–11