To Remove And Replace An Acquisition Hybrid; To Remove The Acquisition Hybrid - Agilent Technologies 54835A Service Manual

Hide thumbs Also See for 54835A:
Table of Contents

Advertisement

Chapter 6: Replacing Assemblies

To remove and replace an acquisition hybrid

To remove and replace an acquisition hybrid

Use the following procedure to remove and replace an acquisition hybrid. When necessary, refer
to other removal procedures.
CA UT IO N
ELECTROSTATIC DISCHARGE!
Use grounded wrist straps and mats when servicing the acquisition board. Electrostatic discharge
can damage electronic components.
You do not have to remove the acquisition board before replacing an acquisition hybrid.
To understand the sequence of parts, use the accompanying illustration.

To remove the acquisition hybrid

Disconnect the power cable and remove the cover.
1
Use a T-6 Torx driver to remove two screws that secure the heatsink spring, then
2
remove the heatsink.
Disconnect the semi-rigid cables from the A/D-FISO hybrids, then disconnect from the
3
attenuator ends.
First, use a T-6 Torx driver to remove the four screws that secure the top plate.
4
Use a 3/16 hex driver to remove two standoffs that secure the top plate.
5
Remove the top plate.
6
Grasp the hybrid across the diagonal corners (or sides) using your thumb and index
7
finger. Then lift the hybrid off of the connector assembly.
6–32