Agilent Technologies 54835A Service Manual page 119

Hide thumbs Also See for 54835A:
Table of Contents

Advertisement

Re-enter the WINBIOS to change one parameter that could not be set in step 6.
7
a Press F1 to enter the BIOS Setup.
b Double-click on Detect IDE Utility. The line labeled "Pri. Master" should contain numeric
information indicating that it has found a hard drive. Exit this utility.
If a hard drive was not found, check Motherboard Hard Drive Cable W9 and SVGA to Hard
Drive Cable W10. If the drive is still not found, replace the hard disk drive.
c Exit the BIOS Setup and save the changes you made.
Table 5-8 summarizes the settings for the system configuration.
Table 5-8
Series 757 System BIOS Settings with 1.44 Mbyte floppy drive
Configuration Item
Load Optimal Defaults
Standard Setup—Date/Time
Standard Setup—Floppy Drive A
Standard Setup—Floppy Drive B
Advanced Setup—System Keyboard
PCI/PnP Setup—IRQ 9
Peripheral Setup (OnBoard Primary/Secondary IDE)
Detect IDE Utility (Hard drive Primary Master)
Power Management/APM*
Hard Disk Power Down Mode*
Hard Disk Time Out (minute)*
* These configurations are only for instruments with 64 Mb or more of RAM.
The oscilloscope should now reboot, load Windows 95, and come up with the scope display. If it
does not, recheck your steps in the above procedure. If all the settings are correct, return to the
Primary Troubleshooting Procedure.
To configure the motherboard jumpers and set up the BIOS
Setting
YES
Enter current values
1.44 MB, 3-1/2
Not Installed
Absent
ISA
Primary
Data
Enabled
Suspend
15 min
Chapter 5: Troubleshooting
5–41