To Remove And Replace The Cover - Agilent Technologies 54835A Service Manual

Hide thumbs Also See for 54835A:
Table of Contents

Advertisement

Chapter 6: Replacing Assemblies

To remove and replace the cover

To remove and replace the cover

Use these steps to remove and replace the cover. When necessary, refer to other removal
procedures.
Disconnect the power cable.
1
Disconnect all scope probes and BNC input cables from the front panel.
2
Disconnect any other cables, such as mouse, keyboard, printer, or GPIB cables.
3
Remove the two Torx T15 screws securing the side handle.
4
Remove the four Torx T15 screws that secure the rear feet (two in each foot).
5
Remove the four Torx T15 screws that secure the cover to the chassis.
6
Set the unit on its side. Carefully slide the cabinet off of the frame by pulling the front
7
panel and the cover away from each other.
To replace the cover, reverse the above procedure.
8
Be sure to keep ribbon cables out of the way when replacing the cover, particularly the flex cable
and connector for the AutoProbe assembly at the bottom front of the oscilloscope.
CA UT IO N
PROPERLY TIGHTEN HANDLE AND SCREWS!
Tighten the side handle and rear feet screws to 2 Nm (18 in-lbs).
Figure 6-1
Cover Fasteners
6–4

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

54845a54846a

Table of Contents