c From the back of the front panel, use a small screw driver or other slender pointed object
through the access hole to push the AutoProbe assembly faceplate away from the front
panel assembly.
Figure 6-6
Pushing Out the AutoProbe Faceplate
CA UT IO N
AVOID DAMAGE TO THE RIBBON CABLE AND FACEPLATE!
Do not pry around the edge of the assembly. Doing so may damage the ribbon cable or faceplate.
d Finish removing the AutoProbe assembly by feeding the mylar ribbon cable through the
front panel slot.
To replace the AutoProbe assembly, reverse the above procedure.
3
Chapter 6: Replacing Assemblies
To remove and replace the AutoProbe assembly
6–7