Maintenance; Cleaning Instructions - Barco Eonis MDRC-2122 WP User Manual

22–inch clinical display
Hide thumbs Also See for Eonis MDRC-2122 WP:
Table of Contents

Advertisement

6. MAINTENANCE

General maintenance information
The Eonis display does not require any scheduled maintenance or calibration activities. We recommend
to use QAWeb with the Barco default tests and frequencies to calibrate and maintain the display or return
the display to a Barco approved maintenance organization. In any case of doubts, contact the Barco
Healthcare Division.
6.1

Cleaning instructions

To clean the display
Clean the display using a sponge, cleaning cloth or soft tissue, lightly moistened with a recognized clean-
ing product for medical equipment. Read and follow all label instructions on the cleaning product. In case
of doubt about a certain cleaning product, use plain water.
Do not use following products:
For black model; alcohol/solvents at higher concentration > 5%, for white model; alcohol/solvents at
higher concentration > 70%
Strong alkalis lye, strong solvents
Acid
Detergents with fluoride
Detergents with ammonia
Detergents with abrasives
Steel wool
Sponge with abrasives
Steel blades
Cloth with steel thread
C
AUTION
rings or other jewelry and do not apply excessive pressure on the front glass or LCD.
C
AUTION
excess liquid may cause damage to internal electronics. Instead, apply the liquid to a
cleaning cloth.
K5903049 EONIS 25/06/2015
:
Take care not to damage or scratch the front glass or LCD. Be careful with
:
For black model only: Do not apply or spray liquid directly to the display as
6. Maintenance
21

Advertisement

Table of Contents
loading

Table of Contents