DETAILED ASSEMBLY
Fig. 6
Fig. 7
Fig. 8
Fig. 9
ACCU-SCOPE
®
(continued)
73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com
EXC-500 MICROSCOPE SERIES
Installing the Nosepiece - (Fig. 6)
Unscrew the dust plugs from the nosepiece ① for the number
of objectives you will be installing.
Using the Hex wrench, loosen the lock screw ② enough so
that the nosepiece can easily slide into the notch ③ on the
front of the dovetail mount ④, then slide the nosepiece onto
the mount.
Retighten the lock screw ② with the Hex wrench.
Installing the Viewing Head - (Fig. 7)
Using the Hex wrench loosen the lock screw ①.
Position the viewing head ② above the dovetail opening ③
as shown and with it tilted slightly down on the right, slide the
dovetail under the notches in the dovetail hole and set it into
place with the two eyepiece tubes ④ facing forward.
Retighten the lock screw ①.
Installing the Eyepieces - (Fig. 8 & 9)
Carefully remove the eyepieces ① from the protective
packaging – be sure not to touch any optical (glass)
surfaces. Remove the dust caps.
Insert an eyepiece ① into one of the eyepiece tubes ② and
gently twist and push the eyepiece in until it is flush with the
top surface of the eyepiece tube.
Repeat above for the other eyepiece.