6
Performance Tests and Calibration
Step
Test Function
11
Press
µA mode
12
Turn the Rotary Switch to the
Move the test lead from uA.mA and COM terminal to A and COM terminal
Caution: Connect the calibrator to hanheld multimeter's A and COM terminals before applying 3A and 10A
Move the test lead from A and COM terminal to uA.mA and COM terminal
13
Press
mA mode
Move the test lead from uA.mA and COM terminal to A and COM terminal
Caution: Connect the calibrator to hanheld multimeter's A and COM terminals before applying 3A and 10A
14
Press
A mode
[1]
Be sure to recalibrate "Short" using the dual banana plug with copper wire after performing calibration for resistance.
138
Cal Range
button to go to
500 µA
5000 µA
Open
position
50 mA
440 mA
5 A
10 A
button to go to
50 mA
440 mA
button to go to
5 A
10 A
Input
U1251B
30 µA,1 kHz
030.00 µA
300 µA,1 kHz
300.00 uA
300 µA,1 kHz
0300.0 µA
3000 µA, 1 kHz
3000.0 µA
Input terminal
oPEn
open (remove any
test leads and
Short Plugs from
the input terminal)
30 mA
30.000 mA
300 mA
300.00 mA
3 A
3.0000 A
10 A
10.000 A
3 mA, 1 kHz
03.000 mA
30 mA, 1 kHz
30.000 mA
30 mA, 1 kHz
030.00 mA
300 mA, 1 kHz
300.00 mA
0.3 A, 1 kHz
0.3000 A
3 A, 1 kHz
3.0000 A
3 A, 1 kHz
3.0000 A
10 A, 1 kHz
10.000 A
Agilent U1251B/U1252B User's and Service Guide
Cal Item
U1252B
030.00 µA
300.00 µA
0300.0 µA
3000.0 µA
oPEn
30.000 mA
300.00 mA
3.0000 A
10.000 A
03.000 mA
30.000 mA
030.00 mA
300.00 mA
0.3000 A
3.0000 A
3.0000 A
10.000 A