Valid Adjustment Input Values

Agilent U1251B/U1252B User's and Service Guide
Adjustment can be accomplished using the following input
values below.
Table 6-3 Valid adjustment input values
Function
Range
5V, 50 V, 500 V, 1000 V
5 V, 50 V, 500 V, 1000 V
(for U1251B)
5 V, 50 V ,500 V, 1000 V
(for U1252B)
50 mV, 500 mV, 1000 mV
500
50 mA, 440 mA, 5 A, 10 A
Ω
500 Ω, 5kΩ, 50 kΩ,500 kΩ, 5MΩ,
50 MΩ
Diode
/
10 nF, 100 nF, 1000 nF, 10
F, 1000
µ
Performance Tests and Calibration
A, 5000
A
µ
µ
F, 100
µ
F, 10 mF / 0 °C
µ
Valid Amplitude Input Values
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
Make sure to provide 0 °C
with ambient compensation
6
131

Advertisement