U1252B Ac+Dc Specifications - Agilent Technologies U1251B User's And Service Manual

Handheld digital multimeter
Hide thumbs Also See for U1251B:
Table of Contents

Advertisement

7
Specifications

U1252B AC+DC Specifications

Table 7-4 U1252B AC Accuracy ± (% of reading + No. of Least Significant Digit)
Function
Range
[2]
50.000 mV
Voltage
500.00 mV
1000.0 mV
5.0000 V
50.000 V
500.00 V
1000.0 V
Function
Range
500.00 μA
Current
5000.0 μA
50.000 mA
440.00 mA
5.0000 A
10.000 A
146
30 Hz –
Resolution
45 Hz
0.001 mV
1.5+80
0.01 mV
1.5+65
0.1 mV
1.5+65
0.0001 V
1.5+65
0.001 V
1.5+65
0.01 V
1.5+65
0.1 V
1.5+65
Resolution
[4]
0.01 μA
0.1 μA
0.001 mA
0.01 mA
0.0001 A
(5)
0.001 A
[1] The additional error to be added as frequency >20 kHz and signal input <10
%of range: 3 counts of LSD per kHz.
[2] Input impedance: Refer to
[3] The input signal is lower than the product of 20,000,000 V-Hz (product of
voltage and frequency).
[4] Input current >35 μArms.
[5] Current can be measured from 2.5 A up to 10 A continuously. Add 0.5% to
the specified accuracy if the signal measured is in the range of 10 A to 20 A and
for a period of up to 30 seconds. After measuring current >10 A, leave the meter
to cool down for a period that is twice the measuring time used before
application of low current measurement.
[6] Input current < 3 Arms.
Frequency
45 Hz –
1 kHz –
1 kHz
10 kHz
0.4+60
0.7+60
0.4+30
0.4+30
0.4+30
0.4+30
0.4+30
0.6+30
0.4+30
0.4+30
0.4+30
0.4+30
0.4+45
0.4+45
Frequency
30 Hz – 45 Hz
45 Hz – 1 kHz
1.1+25
1.1+25
1.2+25
1.2+25
[6]
1.8+30
[6]
1.8+30
Table
7-6.
Agilent U1251B/U1252B User's and Service Guide
10 kHz –
20 kHz
0.8+60
0.8+45
0.8+45
1.5+45
1.5+45
1.5+45
[3]
1.5+45
1 kHz – 20 kHz
0.8+25
0.8+25
0.8+25
0.8+25
0.9+25
0.9+25
0.9+25
0.9+25
0.9+30
3.3+70
<3 A/5 kHz
0.9+25
20kHz –
[1]
100kHz
3.5+220
3.5+125
3.5+125
3.5+125
3.5+125
[3]
3.5+125
N/A