Adam Equipment HIGHLAND SERIES Manual page 21

Hide thumbs Also See for HIGHLAND SERIES:
Table of Contents

Advertisement

5) The  display  will  show  the  first  weight  that you  can use  to calibrate the  unit. 
You can change this value by pressing the [Mode]  key. The weights that can 
be used are as follows:  
Model # 
HCB123  HCB153  HCB302  HCB602  HCB602H
Weight 1 
60g 
Weight 2 
120g 
Weight 3 
‐ 
 
6) Once the calibration weight has been selected press the [Tare] key 
8) The display shows LoAd. Place your weight on the pan. Press [Tare] key once 
the stable sign is shown.  
9)  The display will show PASS. Remove the weight from the pan.  
 
NOTE: If the calibration fails retry.  The balance will show FAlL H (when the weight is 
higher)  or  FAlL L  (when  the  weight  is  lower).  Repeat  the  process  using  the  correct 
calibration weight. 
 
21 | P a g e  
 
50g 
100g 
100g 
200g 
150g 
300g 
 
 
 
200g 
200g 
400g 
400g 
600g 
600g 
 
 
© Adam Equipment Company 2008 
HCB1002  HCB1502  HCB3001 
500g 
500g 
1000g 
1000g 
‐ 
1500g 
1000g 
2000g 
3000g 

Advertisement

Table of Contents
loading