Cancelling Doctor's Function
To clear the device again and to reallow access to other programs, press and
-
hold the
button again for approx. 5 seconds. To do this, the impulse in-
must be set to 00. Deletion of the Doctor's Function is
tensity of
and
confirmed with a long acoustic signal.

8. Cleaning and storage

Adhesive electrodes
• To ensure that the adhesive electrodes remain adhesive for as long as pos-
sible, clean them carefully with a damp, lint-free cloth or clean the
underside of the electrodes under lukewarm running water and pat dry with a
lint-free cloth.
 
Before cleaning with water, remove the connection cables from the
electrodes.
• After use, stick the electrodes onto the backing film.
Cleaning the unit
• Remove the batteries from the device every time you clean.
• Clean the unit after use with a soft, slightly moistened cloth. In case of more
extreme soiling you can also moisten the cloth with mild soapy water.
• Do not use any chemical cleaners or abrasive agents for cleaning.
Ensure that no water enters the device.
Storage
• Remove the batteries from the unit if you are not going to use it for a longer
period. Leaking batteries can damage the unit.
• Do not make any sharp kinks in the connecting leads or electrodes.
• Detach the connecting cable from the electrodes.
• After use, stick the electrodes onto the backing film.
• Store the machine in a cool, well-ventilated place.
• Never place any heavy objects on the machine.

9. Disposal

Used, fully discharged batteries must be disposed of in a specially labeled col-
lection container, at toxic waste collection points or through an electrical retailer.
You are under legal obligation to dispose of batteries correctly.
Note: You will find these markings on batteries containing harmful
substances: Pb = battery containing lead, Cd = battery containing
cadmium, Hg = battery containing mercury.
Please dispose of the device in accordance with the directive
2002/96/EC – WEEE (Waste Electrical and Electronic Equipment). If
you have any queries, please refer to the local authorities responsible
for waste disposal.

10. Problems and solutions

The machine does not switch on when the ON/OFF button
What should I do?
(1) Make sure that the batteries have been inserted correctly and have contact.
(2) Change the batteries if necessary.
(3) Contact customer service.
The electrodes are detaching themselves from the body. What should I do?
(1) Clean the adhesive surface of the electrodes with a damp, lint-free cloth.
Then let them dry in the air and attach them again. Should the electrodes still
not adhere firmly, they must be replaced.
(2) Before each application, clean the skin, avoiding using any skin cream or oils
before treatment. Shaving can increase adhesion of the electrodes.
There is no noticeable stimulation. What should I do?
(1) Interrupt the programme by pressing the ON/OFF button
the connecting leads are properly connected to the electrodes. Make sure
that there is a firm contact between the electrodes and the treatment area.
(2) Make sure that the plug of the connecting lead is firmly connected to the ma-
chine.
(3) Press the ON/OFF key
to start the programme again.
(4) Check that the electrodes are positioned properly and make sure that the ad-
hesive electrodes are not overlapping.
(5) Increase pulse intensity in stages.
(6) The batteries are almost empty. Replace these.
What to do if the battery symbol is displayed.
Replace all batteries.
You are aware that the electrodes feel uncomfortable. What should I do?
(1) The electrodes are badly positioned. Check positioning and if necessary
reposition the electrodes.
(2) The electrodes are worn. Due to the fact that current distribution can no
longer be guaranteed evenly over the entire surface, these may lead to skin
irritation. You must therefore replace these.
26