Aligning the Centrifuge

The horizontal alignment of the centrifuge must be checked every time after moving it to a different
location.
The supporting structure must be suitable for horizontal setup of the centrifuge.

Transporting the Centrifuge

• Use a forklift to lift the centrifuge.
• The centrifuge can be damaged by impacts.
• Transport the centrifuge upright and if at all possible in packaging.
Note The packaging is an one-way packaging. Assign a logistic company for the transport. Contact
Customer Service.

Mains Connection

1. Turn off the power supply switch located on a small recessed panel in the upper right corner of the front
cabinet panel (press "0").
2. Plug the centrifuge into grounded electrical sockets only.
3. Check whether the cable complies with the safety standards of your country.
4. Make sure that the voltage and frequency correspond to the figures on the rating plate.
5. Establish the connection to the power supply with the connecting cable.
Thermo Scientific
WARNING Always remove the rotor before moving the centrifuge.
2