Technical Data - Thermo Scientific NS-FANT12-CH Quick Setup Manual

Laboratory centrifuge
Table of Contents

Advertisement

Technical Data

The technical data of the RC BIOS is listed in the following table.
Table 1-1. Technical Data RC B
Feature
Environmental conditions
permissible ambient temperature
Overvoltage category
Pollution degree
Heat dissipation
IP
Running time
Maximum speed n
max
Minimum speed n
min
Minimum sample temperature
Maximum RCF value at n
max
Maximum kinetic energy
Noise level at maximum speed
Temperature setting range
Dimensions
Height
Width
Depth
Length power cord
Weight without rotor
a
front side measurement
Thermo Scientific
IOS
a
Value
-Use in interior spaces
-Altitudes of up to 2,000 m above sea level
-max. relative humidity 80% up to 31 °C; decreasing linearly up to 50%
relative humidity at 40 °C
+2° C to +40° C
+35.6° F to +104° F
II
II
2
2
230V
208V
~5.2 kW
~4.5 kW
17700 BTU/h
15300 BTU/h
20
20
unlimited
unlimited
8500 rpm
8500 rpm
300 rpm
300 rpm
4°C at 8500 rpm
4°C at 7800 rpm
15860 x g
15860 x g
<212 kJ
< 212 kJ
< 65 dB (A)
< 65 dB (A)
-10 °C to +40 °C
-10 °C to +40 °C
1178 mm
46.3 inch
800 mm
31.5 inch
905 mm
35.6 inch
3000 mm
118.1 inch
475 kg
1047 lb
1
Introduction and Description
Technical Data
II
2
400V
~5.8 kW
19700 BTU/h
20
unlimited
8500 rpm
300 rpm
4°C at 8500 rpm
15860 x g
< 212 kJ
< 65 dB (A)
-10 °C to +40 °C
RC B
IOS
1-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rc bios

Table of Contents