Mains Supply

The following table contains an overview of the electrical connection data for the RC B
to be taken into consideration when selecting the mains connection socket.
Table 1-4. Electrical connection data of the RC BIOS
Cat.
Mains voltage
75007034
208-220V
75007035
230V
75007036
220V-230V
75007037
230V/ 400V 3

Rotor Selection

RC B
IOS
The rotor that is available for use in the RC B
F8-6x1000y
The technical data of the rotors and the corresponding adapters and reduction sleeves for various
commercially available containers can be found in the corresponding rotor operating manuals.
For more information visit our website at: http://www.thermo.com
Thermo Scientific
Rated cur-
Frequency
rent
60Hz
23.5 A
60Hz
22.5 A
50Hz
22 A
50Hz
20.5 A
is sold commercially in a package solution (unit and rotor included).
Power con-
Equipment
sumption
fuse
4400W
25A & 6A
4500 W
25A & 6A
4300 W
25A & 6A
3800 W
25A & 6A
:
IOS
096-061137
1
Introduction and Description
Mains Supply
. This data is
IOS
Building
Supply Power
fuse
Plugs
30 A
NEMA 6-30P
SinglePhase
30 A
NEMA 6-30P
SinglePhase
32A
IEC60309 CEE 32
A-6h 1P+N+PE
3x 32A
IEC60309 CEE 32
A-6h 3P+N+PE
RC B
IOS
1-5