Preselecting temperature
You can preselect temperatures between -10 °C and +40 °C.
1. Press the TEMP key in order to enter the temperature.
2. Enter the desired temperature using the numeric pad.
The digits show in sequential order.
3. Confirm your entry by pressing ENTER.
Your entry will be automatically confirmed if you do not press any key for 5 seconds.

Centrifugation

Once the rotor has been properly installed, the main switch has been turned on and the door has been
closed, you can start centrifugation
Starting centrifuge program
1. Press the START key on the control panel.
The centrifuge accelerates to the pre-set speed with the time display active.
2. You cannot open the door while the centrifuge is running.
Imbalance indicator
If a load is imbalanced, this will be indicated by the message "ROTOR IMBALANCE - BALANCE
ROTOR".
The run will terminate.
Check the loading and start the centrifuge once again. See the information on proper loading in the
rotor instruction manual. For information on troubleshooting, see section
on
page
Stopping the centrifugation program
Set run time
Usually the run time is preset and you only have to wait until the centrifuge stops automatically when
the preset time limit expires.
As soon as the speed drops to zero, the message "END" will appear in the SPEED field. By pressing the
OPEN key, you can open the lid and remove the centrifuge material.
You can also stop the centrifugation program manually at any time by pressing the STOP key.
ACE
When you select an ACE controll run, the run will determine when the centrifuge has reached the set
ACE value.
Thermo Scientific
7-3.
4
Operation
Centrifugation
"Troubleshooting by user"
RC B
IOS
4-5