8
Specifications and Standards
Specifications
8.1
Nikon Microscope
Model
Optical system
Focus up/down motion
Lamp ratings
Lamp type
Average lamp life
Input ratings
Power cord
Operating conditions
Transport/storage
conditions
External dimensions and
weight (main unit)
ECLIPSE 50i
ECLIPSE 50i
Infinity-corrected CF optical system
Objective:
CF160
Objectives:
Field number 22 (with ergonomic tube/binocular
tube/cytodiagnostic unit), 25 (with trinocular eyepiece tube T/F)
Nosepiece:
Sextuple
Drive system: Manual coarse/fine motion
(calibration markings for fine motion: 1 µm/marking)
Stroke:
2 mm upward, 28 mm downward
With one-touch refocusing mechanism
6V/30W halogen lamp
Halogen lamp (PHILIPS 5761)
100 hours
100-240V AC; 230V (10%, 50/60Hz, 0.9A; 0.5A)
• 100-120V areas:
UL-approved detachable power cord set (3 conductor grounding Type SVT, AWG
18, 3 m long maximum, rated at 125V AC minimum)
• 220-240V areas:
EU/EN-approved 3-conductor power cord set (3 conductor grounding Type
H05VV-F, 3 m long maximum, rated at 250V AC minimum)
Temperature:
Humidity:
Altitude:
Degree of pollution:
Installation:
Electric shock protection class: Class I
Indoor use only
Temperature:
-20 to 60°C
Humidity:
90% RH max. (no condensation)
External dimensions:
Weight:
0 to 40°C
85% RH max. (no condensation)
2000 m max.
Degree 2
Category II
184 (W) x 358 (H) x 383 (D) mm (excluding projections)
Approx. 9 kg