Specifications - Respironics REMstar Plus with C-Flex User Manual

Continuous positive airway pressure
Table of Contents

Advertisement

Specifications

Device Size
Product Use,
Transport, and
Storage
Standards
Compliance
Mode of Operation
AC Power Consumption
DC Power Consumption
Type of Protection
Against Electric Shock
Degree of Protection
Against Electric Shock
Degree of Protection
Against Ingress of Water
Pressure Range
Pressure Stability
Sound Pressure Level
Maximum Flow
Pressure Display Accy.
4
REMstar Plus with C-Flex User Manual
Dimensions:
Weight:
Temperature:
Humidity:
Atmospheric Pressure: 76.7 to 102 kPascals
This device is designed to conform to the following standards:
IEC 60601-1
General Requirements for Safety of Medical Electrical Equipment
IEC/EN 60601-1-2:2001 (2nd Edition)
EN ISO 17510-1
Sleep Apnea Breathing Therapy Devices
Continuous
100 - 240 VAC, 50/60 Hz, 1.0 A max.
12 VDC, 3.0 A max.
Class II Equipment
Type BF Applied Part
IPX0 - Ordinary Equipment
4 to 20 cm H
O (in 1 cm H
2
4 to 20 cm H
O (±1.0 cm H
2
Measured in accordance with EN ISO 17510 @ 6.6, 13.2, & 20 cm H
BPM performed at 23° C (±2° C), 50% RH (±5%), and an atmospheric pressure of 101.54 kPascals.
<30 dB(A)
Measured in accordance with EN ISO 17510 @ 10 cm H
applies to the CPAP device with or without the optional REMstar Heated Humidifier.
35 LPM
Measured in accordance with EN ISO 17510 @ 6.6, 13.2, & 20 cm H
BPM performed at 23° C (±2° C), 50% RH (±5%), and an atmospheric pressure of 101.54 kPascals.
±(0.15 + 4% of actual reading)
9 1/2" x 6 7/8" x 4 1/2" (24 x 17 x 12 cm)
<4.0 lbs. (<1.8 kg)
Operation
41 to 95° F (5 to 35° C)
15 to 95% Non-condensing 15 to 95% Non-condensing
Immunity for Medical Electrical Equipment
O increments)
2
O)
2
O at the patient circuit. This measurement
2
Transport & Storage
-4 to 140° F (-20 to 60° C)
N/A
O @ 500 ml with BPM set to 10, 15, & 20
2
O @ 500 ml with BPM set to 10, 15, & 20
2