Invacare STORM4MAX Service Manual page 25

Table of Contents

Advertisement

®
Invacare
- STORM
Component
Batteries
Remote /
electronics
modules
Drive program
4
MAX
Check
Check batteries for
damage
Check battery voltage
Check battery voltage
Check contacts and
terminals
Remote, status display
blinking
Fixings
Cables and connecting
plugs
Drive lever function
Power supply
Check drive electronics
program version
Remedy
Replace batteries if
necessary
Charge batteries
Charge batteries
Clean contacts and
terminals
Evaluate error/blink
code
Tighten fixings,
replace if necessary
Tighten cables and
connecting plugs,
replace if necessary
Replace drive lever if
necessary
Replace remote if
necessary
Tighten cables and
connecting plugs,
replace if necessary
Update software if
newer version
available.
SERVICE MANUAL
ü
Notes
See chapters 8.4.2
and 8.4.3.
See operating
manual
Please refer to the
safety information
in chapter 8.4.2 and
8.4.3 for handling
batteries
See chapter 6.
See chapter 8.3.