Advanced Entry Calibration (continued)
For example, at Gain Number 8,
these were the sensor readings for
each of the media conditions.
(Actual readings will vary.)
Step
Action
16
Press the
Using the
Key, set the
'Paper' level to the value
determined in the previous
step. Press the
to set the entry (indicated
by the '*') and advance the
menu.
Repeat for the 'Gap' (or
'Mark') and 'Empty'
levels.
17
After all entries have been
made, press the
Key to back out of the
menu and then press the
Key to save the
settings and return to the
Ready Mode.
Note: If the Advanced Entry Calibration fails, try the following procedure:
Re-enter the menu system. Go to Media Settings
Sensor Gain and lower the selected Gain Setting number by one (to make the sensor less sensitive).
Select the new Gain Setting, save the changes, exit the menu, and then test your media. Repeat the
procedure until a usable media setting is obtained.
Adjustments and Maintenance
Displayed Message
Key.


or the


Key


Or, for reflective media:








Paper
Gap (or Mark)
198
084
The selection will flash and the
display heading will change to
indicate the item for entry.
From '
next label TOF.
Calibration
2-9
Empty
014
Comment
', press the
READY
Key to advance to the
Advanced Entry