SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING

Service Indicator Light Diagnostics

NUMBER
ERROR CODE
OF
DESCRIPTION
FLASHES
1
2
3
Left Motor Fault
4
Right Motor Fault
5
Left Park Brake Fault
6
Right Park Brake Fault
7
8
Controller Fault
9
Communications Fault
10
11
Incompatible or incorrect
TDX® SC TDX® Spree
User Fault
Release joystick to neutral and try again.
Battery Fault
Charge the batteries. Refer to Charging Batteries on page 69.
Check that battery cables are connected properly. If necessary,
replace batteries. Refer to Replacing the Batteries on page 65.
Contact Invacare/Dealer for service.
Contact Invacare/Dealer for service.
Contact Invacare/Dealer for service.
Contact Invacare/Dealer for service.
Remote Fault
Check to make sure joystick is connected properly.
Contact Invacare/Dealer for service.
Contact Invacare/Dealer for service.
Contact Invacare/Dealer for service.
General Fault
Contact Invacare/Dealer for service.
Wrong type of remote connected. Contact Invacare/Dealer
Remote
for service.
POSSIBLE SOLUTION