Service Indicator Diagnostics - Invacare ProntoM41 Service Manual

With surestep
Table of Contents

Advertisement

®
Invacare
- PRONTO M41
5.1.3

Service Indicator Diagnostics

Number
of Flashes
1
2
3
4
Diagnostics
Error Code
Code
Description
E01
User Fault
E02
Battery Fault
E03
Left Motor
Fault
E04
Right Motor
Fault
Sub
Details of
Code*
Error Code
00
Stall Timeout or
user error.
00
Recharge batteries
or replace.
00
Left Motor Short
Circuit
01
Left Motor Open
Circuit
02
Left Motor
Connection Fault B-
03
Motor Terminal
Connected to B+
04
Left Motor Voltage
Fault
05
Left Motor Bridge
Fault
06
Too Many
Hardware Current
Limit Events
07
Current Offset Out
of Range
08
Hardware Current
Limit Fault
00
Right Motor Short
Circuit
01
Right Motor Open
Circuit
02
Right Motor
Connection Fault B-
03
Motor Terminal
Connected to B+
04
Right Motor Voltage
Fault
05
Right Motor Bridge
Fault
06
Too Many
Hardware Current
Limit Events
07
Current Offset Out
of Range
SERVICE MANUAL
Possible
Solution
Release joystick to
neutral and try again.
Check the batteries
and cable. Try
charging the
batteries. Batteries
may require
replacing.
Check the left motor,
connections and
motor cable.
Check the right
motor, connections
and motor cable.
17

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pronto m41

Table of Contents