Service Indicator Light Diagnostics - Invacare Arrow RWD Service Manual

Table of Contents

Advertisement

Service Indicator Light Diagnostics

NUMBER
ERROR CODE
OF
DESCRIPTION
FLASHES
1
2
3
Left Motor Fault
4
Right Motor Fault
5
Left Park Brake Fault
6
Right Park Brake Fault
7
8
Controller Fault
9
Communications Fault
10
11
Incompatible or incorrect
MPJ™+, PSR+, PSF+ Joysticks or Displays
SYMPTOM
SPM L Park Brake Fault or
SPM R Park Brake Fault displays
and wheelchair does not drive.
CHARGER PLUGGED IN displays.
SPM Battery Fault displays and the
wheelchair does not drive.
JOYSTICK TIMEOUT displays and
the wheelchair does not drive.
JS REV TOO LARGE
JS FWD TOO LARGE
JS LFT TOO LARGE or
JS RGT TOO LARGE
displays and the wheelchair does not
drive.
Part No. 1104849
User Fault
Release joystick to neutral and try again.
Battery Fault
Charge the batteries. Refer to Charging Batteries on page 57.
Check that battery cables are connected properly. Refer to
Disconnecting/Connecting Battery Cables on page 52.
If necessary, replace batteries.Refer to Replacing Batteries on
page 59.
Contact Invacare/Dealer for service.
Contact Invacare/Dealer for service.
Contact Invacare/Dealer for service.
Contact Invacare/Dealer for service.
Remote Fault
Check to make sure joystick is connected properly.
Contact Invacare/Dealer for service.
Contact Invacare/Dealer for service.
Contact Invacare/Dealer for service.
General Fault
Contact Invacare/Dealer for service.
Wrong type of remote connected. Contact Invacare/Dealer
Remote
for service.
PROBABLE CAUSE
Motor lock levers disengaged
(Error code E9 or E10).
Battery charger connected
(Error code E28).
Batteries need to be charged
(Error code E14).
Joystick or input device is
disconnected (Error code 32).
The joystick or input device is
sending a value outside of the
reverse, forward, left or right
limits (Error codes E01, E02,
E03 or E04).
SECTION 6—TROUBLESHOOTING
POSSIBLE SOLUTION
Engage motor lock levers. Refer to
Disengaging/Engaging Motor Lock
Levers on page 67.
Unplug battery charger from the
wheelchair. Refer to Charging Batteries
on page 57.
Charge batteries. Refer to Charging
Batteries on page 57. If batteries fail to
charge properly, check battery charger
or replace batteries. Refer to Replacing
Batteries on page 59.
Turn off power, reconnect the joystick
of input device and turn power on.
Replace joystick or input device.
51
SOLUTIONS
3G Storm Series® Wheelchairs

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Torque sp rwdRanger x rwd3g storm series

Table of Contents