Load Cell Wire To Indicator Connections - Adam Equipment MDW Service Manual

Table of Contents

Advertisement

Calibration of the new load cell will be required and a check for Repeatability,
Off centre loading and Linearity should be made.
Main PCB faults are very rare, if you have a problem where power to the PCB
is correct and the load cell has been checked and confirmed as fully functional
then replacement of the Main PCB is the most effective method of repair.
If this is the case, the PCB number and model of machine should be made
clear so that the PCB can be programmed and checked first, once checked
the PCB can then sent to be fitted directly to the Indicator.
Calibration can then be performed and the unit should function correctly

6.0 LOAD CELL WIRE TO INDICATOR CONNECTIONS

All load cell to Indicator cables are 4 core and wired via a connector following
the pin numbers as below.
Pin 1 Excitation plus+ Red
Pin 2 Excitation minus – Black
Pin 3 Signal minus – Green
Pin 4 Signal Plus + White
8