Sony CCU-900 Installation And Maintenance Manual page 20

Camera control unit
Hide thumbs Also See for CCU-900:
Table of Contents

Advertisement

1-5. Settings of Switches and Controls on Boards
. S1002-3 (CB CAMERA POWER independent/interlock)
Set whether color bars or gray pattern is output before
the camera signal is output when the power of the CCU
is turned on.
ON : Interlock (Outputs color bar or gray pattern
according to the setting of S1003-3.)
OFF : Independent (Not output.)
Factory setting : OFF
. S1002-4 (MONI SEL system 1, 2 independent/interlock)
Select a control system of the PIX1/2 and WF1/2
connector output.
ON : Interlock (Controls PIX1 and PIX2, and WF1
and WF2 simultaneously from MSU or RCP.)
OFF : Independent (Controls PIX1 and WF1 from
RCP and PIX2 and WF2 from MSU.)
Factory setting : OFF
. S1002-5 (board diagnosis display ON/OFF)
When the CHARACTER button on the control panel of
RCP goes off, set whether the error message is displayed
during Auto Setup and Diagnosis (self-diagnosis).
ON : Does not display the error message.
OFF : Normal mode (Displays the error message.)
Factory setting : OFF
. S1002-6
Not used.
Factory setting : OFF
. S1002-7 (CB CHARACTER OFF/ON)
Set ON or OFF of the character signal that is super-
imposed on the color-bars signal output from CCU.
ON : Character superimposed
OFF : Character not superimposed
Factory setting: OFF
. S1002-8 (BACKUP data initialization)
Use this switch to return all the set values to the factory
settings. Set the switch to ON and turn on the power of
the CCU to return to the factory settings.
n
Be sure to set the switch to OFF after initialization.
Factory setting : OFF
. S1003-1
Not used.
Factory setting : OFF
1-12 (E)
. S1003-2 (No Light Gray)
When the incident light level is lower than the specified
level, gray pattern is output from the CCU.
ON : Function OFF (Gray pattern is not output)
OFF : Function ON (Gray pattern is output)
Factory setting : OFF
. S1003-3 (CB/GRAY during CB interlock)
Select the pattern that is output when S1002-2 is inter-
locked.
ON : Gray pattern
OFF : Color bars
Factory setting: OFF
. S1003-4
Not used.
Factory setting : OFF
. S1003-5 (external TALLY/S-bus)
Select whether the INTERCOM/TALLY/PGM connec-
tor of the rear panel is used for TALLY control or
whether S-bus/CNU is used for TALLY control.
ON : S-bus
OFF : INTERCOM/TALLY/PGM connector
Factory setting : OFF
. S1003-6 (WFM PRESET reflection address)
When S1002-3 is set to OFF, select whether the RE-
CALL signal input from the WF MODE connector (D-
sub 15-pin) is interlocked with RCP or MSU.
ON : MSU
OFF : RCP
Factory setting : All OFF
. S1003-7 to 8
Not used.
Factory setting : All OFF
. S1004 (MENU ON/OFF)
Use this switch to display the setup menu.
ON : Displays.
OFF : Does not display.
. S1006 (MENU ENTER/CANCEL)
When S1004 is set to ON, use this switch to select items
from the setup menu, cancel the setting or return to the
default setting.
CCU-900/900P IMM