1) To set this parameter press the [Mode] key during self-checking test when
turning on the balance.
2) After a few seconds, the display will show the first function F1 UNT.
3) Keep pressing the [Mode] key until F5 IEC is displayed.
4) Press the [Tare] key to view the current settings.
5) To change the settings press the [Mode] key to change from INt (internal
calibration) or E (external calibration)
6) Press [Tare] to store a particular setting. The display will return to F5 IEC
7) When F5 IEC is displayed press the [Print] key to return to weighing or press
[Mode] to go to the next function.
6.6

ADJUST THE VALUE OF THE INTERNAL MASS

The internal mass value stored in memory can be adjusted to more closely match
the value of the user's external mass.
1) To set this parameter press the [Mode] key during self-checking test when
turning on the balance.
2) After a few seconds, the display will show the first function F1 UNT.
3) Keep pressing the [Mode] key until F6 CA is displayed.
4) Press the [Tare] key to view the current settings.
5) The display will show the current value with the first digit flashing. To change
the value press [%] to shift the position, press [Mode] to increase the value
and press [Print] to decrease the value. The internal calibration weight should
only change between 95.000 and 105.000 grams or 495.00 and 505.00 grams
(depending upon the model).
6) Press [Tare] to store a particular setting. The display will return to F6 CA
7) When F6 CA is displayed press the [Print] key to return to weighing or press
[Mode] to go to the next function.