8. Turn the concentrator Off (O).
9. Unclamp the pressure gauge and remove it from the T fitting.

After Testing

NOTE: For this procedure, refer to FIGURE 25.4.
1. Insert 6 mm (¼-inch) plug and 6 mm x 5 cm (¼ x 2-inch) silicone tubing onto the T
fitting and secure with 6 mm (¼-inch) clamp.
2. Reinstall the cabinet. Refer to Removing Cabinet on page 26.
Part No 1118396
SECTION 25—SIEVE BED PRESSURE CHECK
Adjustable
Clamp
FIGURE 25.3 Pressure Testing
FIGURE 25.4 After Testing
95
6 mm (¼-inch)
Clamp (STEP 1)
Pressure
Gauge
Product
Tank
6 mm (¼-inch) Plug
6 mm (¼-inch)
Clamp
Platinum™Oxygen