After Testing - Invacare Platinum 5 Service Manual

Platinum series 5 liter oxygen concentrator
Hide thumbs Also See for Platinum 5:
Table of Contents

Advertisement

SECTION 25—SIEVE BED PRESSURE CHECK
7. If pressure in the concentrator does not fall into the specifications listed above, refer
to Troubleshooting on page 17.
8. Turn the concentrator off (O).
9. Unclamp the pressure gauge and remove it from the "T" fitting.

After Testing

NOTE: For this procedure, refer to FIGURE 25.4.
1. Insert ¼-inch plug and ¼ x 2-inch silicone tubing onto the "T" fitting and secure with
¼-inch clamp.
2. Reinstall the cabinet. Refer to Removing Cabinet on page 25.
Platinum™Series
Adjustable
Clamp
FIGURE 25.3 Pressure Testing
FIGURE 25.4 After Testing
98
¼-inch Clamp
(STEP 1)
Pressure
Gauge
Product
Tank
¼-inch Plug
¼-inch Clamp
Part No 1110538