SECTION 20—4-WAY VALVE
NOTE: The Numbers indicate the mounting
screw torque sequence: pre-torque to 1.1 N-M ±
0.2 N-M (10 in-lbs ± 2 in-lbs), then torque to 2.5
N-M ± 0.2 N-M (22 in-lbs ± 2 in-lbs).
FIGURE 20.1 Replacing 4-Way Valve - Manufactured Before 8/1/05
4-Way Valve Face
NOTE: The Numbers indicate the mounting
screw torque sequence: pre-torque to 1.1 N-M ±
0.2 N-M (10 in-lbs ± 2 in-lbs), then torque to 2.5
N-M ± 0.2 N-M (22 in-lbs ± 2 in-lbs).
FIGURE 20.2 Replacing 4-Way Valve - Manufactured After 8/1/05
Platinum™Oxygen
4-Way Valve Face
4-Way Valve
Mounting
Screws
4-Way Valve
Mounting
Screws
70
Manifold
Assembly
Manifold
Assembly
Part No 1118396