Invacare Platinum IRC5LXAW Service Manual page 43

Invacare platinum oxygen concentrator
Hide thumbs Also See for Platinum IRC5LXAW:
Table of Contents

Advertisement

FIGURE 7.6 Removing Compressor and Compressor Heads - Rebuilding the Compressor -
Exhaust Valve Flapper, Valve Restraint and Valve Keeper Strip
NOTE: For this procedure, refer to FIGURE 7.7.
1. Remove the exhaust valve flapper, valve restraint and valve keeper strip from the top
of the valve plate. Discard all.
2. Clean the top of the plate with a clean, soft cloth.
Gasket
NOTE: For this procedure, refer to FIGURE 7.7.
1. Remove the old gasket from the top of the valve plate and discard.
2. Clean the valve plate with a clean, soft cloth.
FIGURE 7.7 Removing Compressor and Compressor Heads - Rebuilding the Compressor -
Exhaust Valve Flapper, Valve Restraint and Valve Keeper Strip
Part No 1118396
O-ring, Intake Valve Keeper and Valve Keeper Strip
Gasket
SECTION 7—COMPRESSOR
Valve Flapper
Screw
Valve
Keeper Strip
O-Ring
Intake Valve
Flapper
Valve Flapper
Screw
Valve
Keeper Strip
Exhaust Valve
Flapper
Valve
Restraint
43
Platinum™Oxygen