Invacare Platinum IRC5LXAW Service Manual page 60

Invacare platinum oxygen concentrator
Hide thumbs Also See for Platinum IRC5LXAW:
Table of Contents

Advertisement

SECTION 13—CONTROL PANEL
11. Remove spade connectors from circuit breaker, hour meter and On/Off (I/O) switch.
(Jumper wire harness from On/Off switch to circuit breaker does not have to be
removed).
12. Remove control panel and reverse STEPS 3-10 to re-install control panel.
13. Run unit and check for leaks at pressure sensor tubing on product tank cap. Refer to
Leak Test on page 85
14. Reinstall cabinet. Refer to Removing Cabinet on page 26.
Hour Meter
(STEP 10)
1/8" I.D. Pressure
Sensor Tubing (STEP 5)
Oxygen Sensor Tube
(SensO
Models Only)
2
Transformer
Harness
Connector
(STEP 10)
Nine or Ten
Pin
Connector
(STEP 8)
8-Pin
Connector
(STEP 9)
Platinum™Oxygen
Circuit Breaker
(STEP 10)
Lower Mounting Screws
FIGURE 13.1 Removing Control Panel
60
Upper Mounting
Screws
ON/OFF
Switch (I/O)
(STEP 11)
Patient Outlet
Barbed fitting
Top and Bottom
Barbed fittings
Control Panel
Part No 1118396