TYPICAL PRODUCT PARAMETERS

Electrical Requirements:
Rated Current Input:
Sound Level:
Altitude:
*Oxygen Output Concentration Levels:
All 5LXO2AWQ/5LXAW/5LXO2AW models
* (Stated concentration levels achieved after ini-
tial warm-up period (approximately 30 minutes).
Maximum Outlet Pressure:
Flow Range:
Low Flow Alarm:
Average Power Consumption:
Pressure Relief Mechanism Operational at:
Change in maximum recommended flow when
back pressure of 7k Pa is applied:
Filters:
Safety System:
Width:
Height:
Depth:
Weight:
Shipping Weight:
Operating Ambient Temperature:
Exhaust:
Oxygen Output:
Cabinet:
Part No 1118396
TYPICAL PRODUCT PARAMETERS
230 VAC ± 10% (253 VAC/207 VAC), 50 Hz
1.4 A (IRC5LXO2AWQ)
2.0 A (IRC5LXAW, IRC5LXO2AW)
39.5 dB Average (IRC5LXO2AWQ)
44 dB Average (IRC5LXAW, IRC5LXO2AW)
Up to 1828 meters (6,000 ft.) above sea level without degrada-
tion of concentration levels. NOT RECOMMENDED FOR USE
ABOVE 1828 M (6,000 ft)
93% minimum at 1 to 3 L/min.
91% minimum at 4 L/min.
87% minimum at 5 L/min (maximum recommended flow)
34.5 ± 3.45 kPa (5 ± 0.5 psi)
0.5 to 5 L/min. Flowrates less than 1 L/min. are not recommend
0 L/min. to 0.5 L/min. Rapid Audible Alarm Beeping (No Acces-
sories Connected).
320 W (IRC5LXO2AWQ)
395 W (IRC5LXO2AW, IRC5LXAW)
241 kPa ± 24.1 kPa (35 psi ± 3.5 psi)
0.7 L/min.
Cabinet (2), Outlet HEPA and Compressor Inlet.
Current overload or line surge shutdown.
High temperature compressor shutdown.
High Pressure Alarm w/compressor shutdown.
Low Pressure Alarm w/compressor shutdown.
Battery Free Power Loss Alarm.
SensO
Oxygen System (SensO
2
46.7 cm ± 1 cm (18 3/8 in ± 3/8 in)
67.0 cm ± 1 cm (26 3/8 in. ± 3/8 in)
36.5 cm ± 1 cm (14 3/8 in. ± 3/8 in)
23.6 kg ± 1 kg (52 lbs ± 2 lbs)
26.8 kg ± 1 kg (59 lbs ± 2 lbs)
10°C - 35°C (50°F - 95°F) at 20-60% relative humidity.
Less than Ambient +19°C (+ 35°F)
Less than Ambient (+ 4°F) +2°C.
Impact Resistant flame-retardant plastic cabinet that conforms
to UL 94-V0.
7
Model) Low Flow Alarm.
2
Platinum™Oxygen