FIGURE 7.4 Removing Compressor and Compressor Heads - Rebuilding the Compressor -
Fan
NOTE: For this procedure, refer to FIGURE 7.5 on page 42.
1. Remove the fan by pulling it straight off the motor shaft grabbing the center housing.
DO NOT pull the fan blades.
NOTE: Note position/orientation of fan for reassembly.
Connecting Rod and Eccentric Assembly
NOTE: For this procedure, refer to FIGURE 7.5 on page 42.
1. Rotate shaft to align the eccentric's set screw with the access hole in the bottom of the
compressor housing.
2. Insert the 1/8-inch Allen wrench into the access hole in the bottom of the compressor
housing (Detail "A" of FIGURE 7.5).
3. Loosen the set screw ¼ turn.
4. Rotate connecting rod to top dead center (180°) and slide the connecting rod/eccentric
assembly off the shaft through the opening in the housing.
Part No 1118396
Head Screws
Valve
Plate
Compressor - Compressor Head Plate - Valve Plates
SECTION 7—COMPRESSOR
Head
Screws
Compressor
Head Plate
Valve
Plate
Compressor
41