Safety; Typical Product Parameters - Invacare XPO100 User Manual

Portable oxygen concentrator
Hide thumbs Also See for XPO100:
Table of Contents

Advertisement

Available languages

Available languages

2 Safety

2.1 Typical Product Parameters

Electrical Requirements:
Rated Current Input:
Sound Level at setting 2:
Altitude:
Oxygen Concentration:*
*Based on an atmospheric pressure of 14.7 psi (101 kPa) at 70°F (21°C)
Conserver Trigger Sensitivity:
Factory set — no adjustment
Pressure activated
Conserver Trigger Delay:
Factory set — no adjustment
Conserver Breath Rate Capacity:
Conserver Bolus Delivery:
Fixed minute volume
Dimensions:
1148112-F~12
AC Power Supply: 120VAC 60 Hz and 230VAC 50 Hz
DC Power Supply: 11VDC, 6.5A
3.8 A at 18 VDC
< 42.7dBA weighted
Up to 10,000 ft (3046 m) above sea level
87% to 95.6%, after initial warm-up period (approximately 5 minutes)
< 0.20 cmH
O pressure drop
2
All settings — using 7 ft (2.1 m ) cannula
50 mSec
Nominal value — using 7 ft (2.1 m ) cannula
Up to, and including, 35 BPM without reduction of bolus minute volume
Setting 1: 300 cc
Setting 2: 460 cc
Setting 3: 620 cc
Setting 4: 740 cc
Setting 5: 840 cc
10" high x 7" wide x 4" deep (25.4 cm high x 17.8 cm wide x 10.2
cm wide)
Safety
9

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Xpo100b

Table of Contents