Actuators
Actuators
Removal of the actuator for the seat-
to-floor function.
1. Place the wheelchair on a level surface.
2. Fit the service tool. See pages 30-31.
3. Switch off the main power switch on the
control panel.
4. Switch the main fuse to OFF. See illustration.
WARNING
Take great care when working with the seat and its actuators,
there is always a risk of personal injury caused by crushing
or pinching.
5. Remove the actuator's upper fixing screw. See
illustration.
6. Remove the actuator's lower fixing screw.
See illustration.
7. Unplug the actuator's cable at the contact on
the cabling.
Fitting
Fit in the reverse order.
Use a torque wrench to tighten the screws
holding the actuator in place to 35 ft-lbs (47Nm)
WARNING
The actuator must be calibrated after fitting. Detailed
information on calibration is available in the technical
manual for the ICS control system.
The seat-to-floor function's actuator.
Upper fixing screw of the actuator
Lower fixing screw of the actuator