Requirements For Module-Level Repair And Replacement; Nibp Characterization - Welch Allyn 53 Service Manual

300 series vital signs monitor
Table of Contents

Advertisement

Troubleshooting and Repair

Requirements for Module-level Repair and Replacement

All repairs to a monitor under warranty must be performed at an authorized
Welch Allyn service center.
Activity
Update the software
Replace the main board
Replace the NIBP module.
Replace the SpO
board
2
Replace the SpO
side panel.
2
Replace the temperature board.
Replace the printer.
Replace the front panel.

NIBP Characterization

Warning Whenever any component of the NIBP assembly (pump, air filter,
check valve, tubing, or NIBP board) is replaced, the NIBP module must be
characterized before the monitor is returned to service.
If the NIBP module is not properly characterized, the NIBP cuff could
overinflate, which could be hazardous to neonatal patients.
The Welch Allyn Monitor Service Utility facilitates this characterization procedure. See
page 40
for more information.
Vital Signs Monitor Series 300
Requirements for Module-level Repair and Replacement
Recommended Action
Return the monitor to an authorized service center.
Return the monitor to an authorized service center.
Replacement procedure beginning on
Perform a complete functional verification
Warning: Always characterize NIBP after replacing any
NIBP component. (See
Replacement procedure beginning on
Perform a complete functional verification
Replacement procedure beginning on
Perform a complete functional verification
Replacement procedure beginning on
Perform a complete functional verification
Replacement procedure beginning on
Perform a complete functional verification
Replacement procedure beginning on
Perform a complete functional verification
39
page
43.
(page
17).
page
39.)
page
43.
(page
17).
page
43.
(page
17).
page
43.
(page
17).
page
43.
(page
17).
page
43.
(page
17).