Disassembly Procedure
printer-frame "piano" legs are toward the front, extending downward from the
assembly.
c. Partially insert the printer frame assembly into the printer well in the rear chassis,
aligning the rear side rails for insertion between the two slotted side latches on
the top of the printer well of the rear chassis.
d. Align the bottom of the two printer frame legs toward the top of the vertical
bulkhead slots.
e. Fully insert the printer assembly into the printer well until the side rails engage in
the slotted side latches and the legs click into the vertical slots. Be sure that the
white plastic shield on the housing rests on the top of the SpO
f. Install the printer door by inserting it in the printer slot, tilted slightly forward,
and snapping the latches onto the printer frame spindles.
Vital Signs Monitor Series 300
Remove and Disassemble the Printer Assembly
67
board.
2
Service Manual