Battery; Patient Isolation Test - Welch Allyn 53 Service Manual

300 series vital signs monitor
Table of Contents

Advertisement

Functional Verification Procedure

Battery

1. Disconnect the battery substitution-cable connector.
2. Install and connect the battery.
3. Connect the AC power adapter to the monitor.
4. Verify that the AC mains indicator
The indicator could be flashing or steady, depending on the state of the battery.
flashing indicates that the battery is charging;
charged to at least 90% capacity.
5. Charge the battery for a minimum of 12 hours (until
6. Turn the monitor on and set it up as follows:
NIBP Auto Interval
Printer, if present
SpO
sensor
2
(if SpO
option is present)
2
Temperature probe
(if temperature option is present)
7. Note the time and let the monitor run until low battery (
(Run time for a new, fully charged battery is at least 12 hours.)
8. Charge the battery for a minimum of 12 hours (until

Patient Isolation Test

Patient Isolation Test - Overview
To verify proper patient isolation, it is important to run this test following any procedure in
which the monitor is opened.
Warning Failure to run the dielectric test when indicated could cause serious
injury to patients, and could lead to damage to the monitor.
The patient isolation test requires an AC Withstand Voltage (hi-pot) Tester, such as the AR
3605 or equivalent. If this equipment is not available, Welch Allyn can perform the patient
isolation test for you quickly, for a nominal fee.
Service Manual
is illuminated on the monitor front panel.
steady indicates that the battery is
stops flashing).
OFF
OFF (not printing)
Disconnected
Either connected or disconnected
stops flashing).
32
Functional Verification
flashing) is indicated.
Vital Signs Monitor Series 300