Checklist and Test Results Report Form
Here is an example of verification results for a typical monitor.
Test
Battery Charging
Power-on Self-test
Beeper
Monitor Off Current
Init/Idle Current
Baseline Current
Battery Voltage
NIBP Characterization Test
NIBP Verification
Pressure Leakage
Pressure Accuracy
Valve/Pump Current
Inflation
Dump
Printer Current
SpO
2
Temperature
Nurse Call Relay
Battery Charge Time
Battery Discharge Time
Monitor Serial #: 9786756453
Tested by: RJ
Test date: 12/18/2003
Service Manual
Result
charge symbol lit
normal
audible
(mA): 0.00
(mA): 0.467 mA
(mA): 0.199 mA
Monitor (V): 5.94 V
C03
Before (mmHg): 100
80 mmHg SYS: 73
150 mmHg SYS: 142
300 mmHg SYS: 290
Valve/Pump (mA)
Baseline (mA)
Difference (mA)
Elapsed time (sec): 6
Manometer reading (mmHg): 0 (3 sec)
Peak (mA): 0.940 mA
Saturation (%): 90
° Fahrenheit: 105.9
pins 1-2
Alarm (Ω) > 1M No Alarm (Ω) < 1
pins 2-3
Alarm (Ω) < 1
12.5 h
19 h
36
Functional Verification
DMM (V): 5.95 V
After (mmHg): 99
Manometer: 72
Manometer:140
Manometer: 286
0.560
0.199
0.361
Pulse (bpm): 200
° Celsius:
No Alarm (Ω) > 1M
Vital Signs Monitor Series 300
Pass
Fail
(or
N/A)