Equipment Required - Welch Allyn 53 Service Manual

300 series vital signs monitor
Table of Contents

Advertisement

Equipment Required

Equipment Required
The following equipment is required for functional verification of a fully configured monitor.
Commercially Available General-purpose/Medical Test Equipment
Item
Power supply
Digital pressure meter
AC withstand voltage (hi-pot) tester
SpO
functional tester
2
(for testing the monitor only)
SpO
extension cable (required for SRC-MAX)
2
SpO
simulator
2
(for testing the monitor and the SpO
Syringe, 60 mL, Slip tip, Luer
Hi-pot cable connectors
Timer
Temperature test key
(for testing the monitor only)
9600 Temperature calibration tester
(for testing the monitor and the temperature probe)
Neonatal cuff hose, 96"
Neonatal #1 cuff, disposable, box of 10
500 cc air volume
DC power adapter
Battery substitution connector
(Use the female end from the cable set)
Welch Allyn Monitor Service Utility
(required for NIBP repair or replacement;
not required for functional testing)
Service Serial Cable (for use with the Welch Allyn
Monitor Service Utility 840-0676-XX)
NIBP tubing connector, threaded
tubing, 1'
Tee, plastic
Service Manual
sensor)
2
Welch Allyn Accessories and Test Equipment
18
Manufacturer Part Number/Specification
Variable, 0-8 VDC, 0.75A (minimum), with
voltage and current indicators (for 1mA current
measurement)
Netech Digimano 1000 or equivalent
Associated Research 3605 or equivalent
Nellcor SRC-MAX
Nellcor DEC-8
Fluke (Biotek) Index2 XL/XLFE or equivalent
BD (Becton, Dickinson) 309654 or equivalent
See
page
34.
06138-000
01800-210, 110V
01800-500, 220V
01800-810, 220V UK
01800-910, 220V Australia
008-0265-XX
008-0620-XX
T112854
5200-101A
660-0237-XX
840-0676-XX
008-0842-XX
600-0021-XX
600-0179-XX
600-0043-XX
Vital Signs Monitor Series 300
Functional Verification

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents