Lake Shore Model 330 Autotuning Temperature Controller User's Manual
5.8

ERROR MESSAGES

On power up, the Model 330 checks internal memory. There are two potential error messages. The first error,
Er01 , indicates an unsuccessful attempt to write and then read the internal non-volatile RAM. This error is not
user-correctable. Consult the factory.
The second error, Er02 , indicates an unsuccessful attempt to read the internal non-volatile RAM for the Model
ID. Sometimes, initializing memory corrects this error. To initialize memory, press and hold both Escape and
Units for about 20 seconds. Release when the power up sequence begins. Perform this operation only under
extreme circumstances; it erases any user-defined curves in memory.
There are two sensor input error messages. If an input signal from the sensor exceeding full scale is applied
to the input leads, OL indicates the overload. If no signal or a wrong polarity signal exists at the input leads,
Er27 indicates a Zero Error for Channel A or Er28 for Channel B. The Model 330 displays dashes " - - - " if the
input switches are improperly set.
Finally, Er30 appears if the measured heater output does not match the predicted output. The controller turns
the heater off. Check heater resistance, test for shorts in the heater wiring, then turn the heater on again. If the
error message returns, consult the factory.
DIP Switch
DIP Switch
Sensor
Sensor
Settings
Settings
Silicon Diode
1
(DT-400
Series)
Platinum
1
RTD
GaAlAs
1
Diode
Thermocouple
1
(compensation
on)
Thermocouple
1
(compensation
o f f )
Improper Switch Setting
*
DIP Switch S1 is for Channel A and S2 is for Channel B.
5-6
Table 5-1. Sensor Input Setup
Display When
Display When
Input Type
Input Type
*
Key is Pressed
Key is Pressed
2
3
2
3
2
3
2
3
2
3
Sensor
Sensor
Input
Input
Units
Units
Range
Range
Volts
0 to 2.5
Ohms
0 to 300
Volts
0 to 6.0
Millivolts
­10 to 10
Millivolts
­15 to 15
If the internal DIP switches are improperly set, the
display will read “Er” when the Input Type
pressed. The normal front panel display will show
dashes “- - - -” to indicate improper DIP switch setting.
Standard
Standard
Curve(s)
Curve(s)
2 ­ Curve10
4 ­ Curve 10 SoftCal™
3 ­ DIN Curve 43760
None
6 ­ AuFe 0.07% vs. Chromel
7 ­ AuFe 0.03% vs. Chromel
8 ­ Type E (chromel-constantan)
9 ­ Type K (chromel-alumel)
10 ­ Type T (copper-constantan)