Lake Shore Model 330 Autotuning Temperature Controller User's Manual
Zone 10
Zone 09
Zone 08
Zone 07
Zone 06
Zone 05
Zone 04
Zone 03
Zone 02
Zone 01
0 K
3.4.5

Power Up (PUP) Configuration

Store a Model 330 Power Up (PUP) configuration to ensure that
it powers up to a user-defined state after power down.
Parameters including heater range, setpoint, gain, reset, units,
and curve number store in non-volatile memory and are
retained even when the line cord is disconnected.
To view PUP status, hold the Enter key for about 5 seconds to
see the display to the right.
"On" updates power up settings when the user changes them
via the front panel or over the remote interface. "On" is the
default PUP. If left on, the instrument powers up in the same
configuration it was powered down in.
"Off" disables updates to the power up memory. The instrument powers up in the configuration it was in when
the power up feature was turned off.
To change PUP status, hold the Enter key and press the t key to toggle the PUP status between On and Off.
3-14
Default Zone Settings
Heater High
P – 100
Heater High
P – 90
P – 80
Heater High
P – 70
Heater High
P – 60
Heater High
P – 50
Heater High
Heater High
P – 40
Heater High
P – 30
Heater Low
P – 20
Heater Low
P – 10
I – 20
D – 00
I – 20
D – 00
I – 20
D – 00
I – 20
D – 00
I – 20
D – 00
I – 20
D – 00
I – 20
D – 00
I – 20
D – 00
I – 20
D – 00
I – 20
D – 00
A
K
K
B
Setpoint:
100 K
Setpoint:
90 K
Setpoint:
80 K
Setpoint:
70 K
Setpoint:
60 K
Setpoint:
50 K
Setpoint:
40 K
Setpoint:
30 K
Setpoint:
20 K
Setpoint:
10 K
0 K
Sample
LOC
Control
OFF