Lakeshore 330 User Manual page 64

Autotuning temperature controller
Table of Contents

Advertisement

Lake Shore Model 330 Autotuning Temperature Controller User's Manual
ECUR
Edit or Add A Data Point In User Curve.
Input:
ECUR XX,Y.YYYYY,ZZZ.Z
Returned: Nothing
Remarks: Enter the point to be added or edited: XX = curve number from 11 to 31, Y.YYYYY = voltage,
and ZZZ.Z = temperature in kelvin. Data points in Curves 00 thru 10 cannot be edited. If the
Model 330 does not recognize either the units value or the temperature value, it assumes you
are entering a new point and places it in the proper ascending position.
Example: If the point to be edited was input as 0.19083,364.0 and should have been 0.19083,365.0,
input this command: ECUR 11,0.19083,365.0[term]. The Model 330 recognizes the
units field and replaces that data point with the new temperature value.
FREE?
Curve Bytes Free Query.
Input:
FREE?
Returned: XXXX (value from 0000 to 3584)
Remarks: Returns the number of curve storage bytes available for new curve entry. New curves require
at least 100 bytes free. A typical 31 point curve requires 176 bytes.
KCUR
Delete User Curve Data Command.
Input:
KCUR XX
Returned: Nothing
Remarks: Deletes all data stored for the User Curve where XX = user curve number 11 thru 31. Curves
00 thru 10 cannot be deleted. Repacks the remaining curves within the NOVRAM.
SCAL
SoftCal Voltage Entry.
Input:
SCAL AA,X.XXXXX,Y.YYYYY,Z.ZZZZZ
Returned: Nothing
Remarks: Stores the SoftCal voltage values at 4.2 K, 77.32 K, and 300 K, where AA = Curve number
from 11 to 31, X.XXXX = 4.2 K voltage, Y.YYYYY = 77.32 K voltage, and Z.ZZZZZ = 300 K
voltage.
Example: SCAL 12,1.6260,1.0205,0.5189[term]
4-22