Calibration - X-Rite EasyTrax Setup And Operation

Table of Contents

Advertisement

Calibration

Regular calibration of the system is important for maintaining accurate and consistent measurements.
Normally, the software application prompts you for a calibration when it is required. Calibration can also
be manually started from the Press Tool screen under the connection icon whenever desired.
Important: The calibration plaque should be cleaned every 30 days or if noticeably dirty to maintain
calibration accuracy. Refer to Cleaning the Calibration Plaque in the Cleaning section for the procedure.
When prompted by the software, slide open the calibration plaque protective cover at the
1.
home position in the table.
Cal plaque
Position the reticule over the targeting area. Make sure the tips of the reticule are in the
2.
center of the targeting circle. The tips of the reticule should also be aligned with the tips of
the upper sights when looking down from the top.
Initiate the calibration procedure in the application. The indicator LED starts to flash orange.
3.
After the calibration is complete, slide the read head away from the calibration plaque and
4.
position the protective cover back in place.
EasyTrax Setup and Operation
White plaque
Protective cover
Targeting circle
16
Tip of reticule
Targeting circle
Tip of upper sight