Dynamic Reserve; Current Inputs; Bandpass Filter - Stanford Research Systems SR530 Manual

Lock-in amplifier
Table of Contents

Advertisement

In the first method, the lock-in uses the A input in a
'quasi-differential' mode. Here, the lock-in detects
the signal as the voltage between the center and
outer conductors of the A input. The lock-in does
not force A's shield to ground, rather it is
connected to the lock-in's ground via a 10Ω
resistor. Because the lock-in must sense the
shield voltage (in order to avoid the large ground
loop noise between the experiment and the lock-
in) any noise pickup on the shield will appear as
noise to the lock-in. For a low impedance source
(as is the case here) the noise picked up by the
shield will also appear on the center conductor.
This is good, because the lock-in's 100 dB CMRR
will reject most of this common mode noise.
However, not all of the noise can be rejected,
especially the high frequency noise, and so the
lock-in may overload on the high sensitivity
ranges.
Quasi-Differential Connection
The second method of connecting the experiment
to the lock-in is called the 'true-differential' mode.
Here, the lock-in uses the difference between the
center conductors of the A & B inputs as the input
signal. Both of the signal sources are shielded
from spurious pick-up.
True-Differential Connection
With either method, it is important to minimize both
the common mode noise and the common mode
signal. Notice that the signal source is held near
ground potential in both cases. A signal which
appears on both the A & B inputs will not be
perfectly cancelled: the common mode rejection
ratio (CMRR) specifies the degree of cancellation.
For low frequencies the CMRR of 100 dB indicates
that the common mode signal is canceled to 1 part
5
in 10
, but the CMRR decreases by about 6
dB/octave (20 dB/Decade) starting at 1KHz. Even
with a CMRR of 10
signal behaves like 100nV differential signal.
There are some additional considerations in
deciding how to operate the lock-in amplifier:
Dynamic Reserve (DR) is the ratio of the largest
noise signal that the lock-in can tolerate before
overload to the full-scale input. Dynamic reserve
is usually expressed in dB. Thus a DR of 60 dB
means that a noise source 1000 times larger than
a full scale input can be present at the input
without affecting the measurement of the signal. A
higher DR results in a degraded output stability
since most of the gain is DC gain after the phase
sensitive detector. In general, the lowest DR
which does not cause an overload should be used.
The Current Input has a 1 kΩ input impedance
and a current gain of 10
from 500 nA down to 100 fA full scale can be
measured. The impedance of the signal source is
the most important factor to consider in deciding
between voltage and current measurements.
For high source impedances, (>1 MΩ), and small
currents use the current input. Its relatively low
impedance greatly reduces the amplitude and
phase errors caused by the cable capacitance-
source impedance time constant. The cable
capacitance should still be kept small to minimize
the high frequency noise gain of the current
preamplifier.
For moderate source impedances, or larger
currents, the voltage input is preferred. A small
value resistor may be used to shunt the source.
The lock-in then measures the voltage across this
resistor. Select the resistor value to keep the
source bias voltage small while providing enough
signal for the lock-in to measure.
The Auto-Tracking Bandpass Filter has a Q of 5
and follows the reference frequency. The
passband is therefore 1/5 of the reference
frequency. The bandpass filter can provide an
additional 20 dB of dynamic reserve for noise
signals at frequencies outside the passband. The
filter also improves the harmonic rejection of the
lock-in. The second harmonic is attenuated an
additional 13dB and higher harmonics are
30
5
, a 10 mV common mode
6
Volts/Amp. Currents

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents