12.3 Balance parameters

ON
OFF
2s
0
F
Selection
F
T
FIS-BA-def-0411
when
"test"
appears
PASS
Sub-selection
0
20
-
-
31
-
33
-
-
-
-
34
-
Symbol
display
0 Verification flat
0
verified
1
non-verified
000 g-Factor, g/kg
Value pre-setting for decades
010 corresponds to 0.10 g/kg
Correction value according to GEO value
table (see chapter 13 Calibration)
1 Digital filter
0 Steady environment, fast weighing time
1 normal environment, medium weighing time
:
9 Unsteady environment,
slow weighing time
15 Resting time for determination of
stabilised weight
Value pre-setting for decades
Value X 17 ms
Selection of the service
parameter menu.
Selection of the balance
parameters
Hidden input of the service
password.
Without the input of a
password, the parameters can
only be displayed but not
changed. Pass code is 0001.
The position is selected using
0
. The figure is entered
using
Adopt and check service
menu