14 Miscellaneous

Disturbances, causes, remedial action

14.1
Continuous automatic check and test routines are carried out. Status and error
messages are the result of these test routines.
Disturbances
108
Soft Error
Activate the
be continued on the weighing terminal
Hard Error
The error cannot be corrected.
Activating the
started
Causes
No power supply or plug
not connected
Pre-load missing or load
plate jamming.
Weighing range exceeded. Unload the scale.
Vibrations at place of
installation
Material to be weighed in
contact with other parts
Weighing material unstable
Material to be weighed not
positioned properly on the
scale. In contact with other
parts.
Error in weighing terminal.
Error in load carrier
F
key. This permits operations to
ON
OFF
key causes a function test to be
Remedial Action
Connect the direct plug-in transformer
unit
Apply the pre-load.
Ensure that load plate has free play.
Eliminate disturbances at the place of
installation.
Adapt parameter settings.
Position the material to be weighed
properly.
Position it in such a manner as to
prevent it from coming into contact with
other parts.
Activate the
Inform KERN quoting the error
number.
Activate the
momentary disturbances.
ON
OFF
key;
ON
OFF
key to remedy