Set Zero Dissolved Oxygen Calibration

Note: In most calibration screens, press f1 (esc) to return to the measurement mode without saving the calibration.
f1 (cal). In dual and multi channel measurement displays, press
or
to highlight DO/RDO – Channel and press f2 (select).
2. Press
or
to highlight Set Zero and press f3 (select).
f3 (start).
Reading is stable. Accept Auto Calibration Value and 0.00.
f3 (cal done) to export the data to the calibration log or press f2 (print) to export the data to the
Thermo Scientific Orion Star A329 Portable pH/ISE/Conductivity/RDO/DO Meter Reference Guide
30