Guidance and manufacturer's declaration - electromagnetic emissions
The BP4GAPO-2M is intended for use in the electromagnetic environment specified below. The customer or the user of the BP4GAPO-2M should assure that
it is used in such an environment.
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations / flicker emissions
IEC 61000-3-3
Compliance
Electromagnetic environment-guidance
The BP4GAPO-2M uses RF energy only for its internal function. Therefore, its
Group 1
RF emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
Group B
The BP4GAPO-2M is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
Not applicable
supply network that suppies buildings used for domestic purposes.
Not applicable
16